SBIR-STTR Award

An Ultra-Low-Cost Adjustable Ambient Air-Oxygen Blender for a Bubble-Continuous Positive Airway Pressure Device to Deliver Oxygen Therapy to Newborns Undergoing Respiratory Distress
Award last edited on: 5/7/20

Sponsored Program
SBIR
Awarding Agency
NIH : NHLBI
Total Award Amount
$149,960
Award Phase
1
Solicitation Topic Code
-----

Principal Investigator
Anuj Bellare

Company Information

Vayu Global Health Innovations LLC

22 Welgate Road
Medford, MA 02155
   (617) 512-3521
   N/A
   www.vayuinnovations.org
Location: Single
Congr. District: 05
County: Middlesex

Phase I

Contract Number: 1R43HL145876-01
Start Date: 3/1/19    Completed: 2/29/20
Phase I year
2019
Phase I Amount
$149,960
Respiratory distress syndrome associated with prematurity, sepsis and birth asphyxia is a major reason for the mortality of newborns worldwide with over three million newborns dying annually and a much larger number disabled, particularly in resource limited regions, worldwide. Oxygen therapy delivered via a continuous positive airway pressure (CPAP) device is the most common treatment to address respiratory compromise in these infants. However, pure oxygen can place the infant at significant risk of oxygen-related damage to the lungs, eyes and brain. A conventional CPAP device is therefore connected to both a compressed air and oxygen tank to control the oxygen content delivered. The conventional CPAP is often prohibitively expensive in low income regions worldwide, requiring uninterrupted electricity, a compressed air tank which isn’t easily available and is insufficiently portable for healthcare workers to treat newborns in remote areas worldwide. A bubble-CPAP (bCPAP) is a simpler, much less expensive device, and has been implemented in resource limited areas of several African and Asian countries. However, improvised bCPAP devices deliver 100% oxygen, thus placing infants at significant risk of organ damage. In this proposal, we hypothesize that a bubble CPAP device equipped with an adjustable Venturi ambient air-oxygen blender will deliver oxygen enriched air efficiently at a controlled percentage ranging from 30 to 80%. As our pilot data on fixed Venturi blenders of different designs shows, the fraction of oxygen delivered depends on their geometry. The Specific Aims are: first, to test the hypothesis that a modular, adjustable Venturi air-oxygen blender can deliver a controlled fraction of oxygen, ranging from 30-80%, in a bCPAP device by designing and fabricating the blender, and measuring the fraction of oxygen delivered; second, to test the hypothesis that the bCPAP device will deliver oxygen efficiently by measuring the minimum flow rate of oxygen required to operate the bCPAP at each fraction of oxygen delivered. Successful completion of this project will allow us to develop a bCPAP device that enables the health care provider the flexibility to dial up the oxygen content as needed without having to change blenders, which would disrupt the treatment and subject the infant to asphyxia in the interim period, thus endangering the infant. Furthermore, this ultra-low cost device requires no electricity and no air tank since it draws in ambient air into the oxygen stream, making it substantially more portable than a conventional CPAP device. Our team has already conducted field-testing and implemented a bubble CPAP package using a fixed Venturi air-oxygen blender in healthcare facilities in Kenya and India, where it is conducting training of local healthcare workers to improve screening, diagnosis and treatment of respiratory conditions of newborns. Our goal is to decrease respiratory newborn mortality by 50% within two years of implementation of our ultra-low cost, bubble CPAP device whose critical component will be our innovative, adjustable ambient air-oxygen blender that we hope to design, fabricate, test and optimize in this proposal.

Public Health Relevance Statement:
Prematurity, sepsis and birth asphyxia result in approximately 3 million deaths among newborns, worldwide, and a much larger number are disabled annually, with respiratory compromise occurring in each of these cases. In this proposal we will design, fabricate and validate an adjustable, handheld, ambient air-oxygen blender, which is a critical component of an ultra-low cost, portable bubble-continuous positive airway pressure device to administer oxygen therapy to newborns undergoing respiratory distress. Our novel ambient air- oxygen blender will, for the first time, allow the healthcare provider to dial the concentration of oxygen delivered to the newborn without disrupting the therapy to change blenders, which would otherwise cause asphyxia, thus endangering the life of the newborn undergoing treatment.

NIH Spending Category:
Assistive Technology; Bioengineering; Infant Mortality; Lung; Neonatal Respiratory Distress; Pediatric; Perinatal Period - Conditions Originating in Perinatal Period; Preterm, Low Birth Weight and Health of the Newborn

Project Terms:
Address; African; Air; Area; Asians; Asphyxia; Asphyxia Neonatorum; base; Brain; Bronchopulmonary Dysplasia; Cessation of life; Clinical; common treatment; Continuous Positive Airway Pressure; cost; Country; Data; design; Device Designs; Devices; Diagnosis; Disabled Persons; Disease; driving force; Electricity; Eye; Fibrosis; field study; flexibility; Gases; Geometry; Goals; Health care facility; Health Personnel; high risk; improved; India; Infant; Inflammation; innovation; Interruption; Kenya; Legal patent; Life; Low income; Lung; Masks; Measures; Morbidity - disease rate; Neonatal Mortality; Newborn Infant; Nose; novel; off-patent; Organ; Oxygen; Oxygen Therapy Care; portability; premature; pressure; Pulmonary Emphysema; Resources; respiratory; Respiratory distress; respiratory distress syndrome; Risk; screening; Sepsis; Stream; Testing; Time; Trainin

Phase II

Contract Number: ----------
Start Date: 00/00/00    Completed: 00/00/00
Phase II year
----
Phase II Amount
----